MITECH MT180 Muti-mode Ultrasonic Thickness Gauge
Mitech MT180 multi-mode ultrasonic thickness gauge, based on ultrasonic measuring principle, it can perform thickness measurements on a wide range of materials including metals, plastic, ceramics glass and other ultrasonic well-conductive materials. It is also capable for measuring the sound velocity of the object with known thickness.
Technical Parameters
Technical Specifications |
Technical Parameters |
Measuring Range |
Support two modes for thickness measurement Pulse-Echo mode: (0.65~600)mm Echo-Echo mode: (3~30)mm |
Accuracy |
±0.05mm(≤10mm);±(0.5%H+0.01)mm(>10mm); H refer to the thickness of workpiece |
Measurement Frequency |
7 times per second for single point measurement, 16 times per second for scan mode measurement. |
Display |
High contrast Segment LCD display (support for EL backlight display) |
Resolution |
0.1mm/0.01mm selectable |
Sound Velocity Range |
(1000~9999)m/s (Capable for measuring the sound velocity of the object with known thickness) |
Probe Calibration |
Zero-point calibration, two-point calibration |
Thickness Measurement Mode |
Single Point measurement, min/max measurement, differential measurement |
Data Storage |
Capable for saving and managing 20 groups of thickness data (up to 100 values for each group) |
Units |
Metric/Imperial unit selectable |
Communication |
USB 1.1 interface for connecting with PC for printing |
Power Source |
With two “AA” size alkaline batteries, it can work above 50 hours continuously with default brightness. |
Auto Power Saving |
Auto shutdown power saving function |
Material |
ABS engineering plastic |
Size |
150mm×74mm×32mm |
Weight |
245g |
Features
l Capable of performing thickness measurements on a wide range of material including metals (such steel, cast iron, aluminum, copper and so on), plastic, ceramics, composites, epoxies, glass and other ultrasonic well-conductive materials.
l With two thickness measurement modes: Pulse- Echo mode and Echo-Echo model, it can measure the thickness through the coating without calculating the coating thickness .
l With large storage capacity and lower power design, it can standby super long time above months.
l Capable for measuring the sound velocity of the object with known thickness, it can meet the thickness testing requirements for various materials.
l Attach with USB data proceeding software, it can connect with PC for data’s analysis, storage and printing.
l Capable for compatible with a variety of probes with different frequency and size, it can identify the probe type automatically.
l With probe-zero calibration and two point calibration functions, it can correct the system error automatically
l Equipped with narrow impulse composite crystal high accuracy probe, it has small dead zone and accurate measurement.
l With EL backlight display, it is convenient for using in dim light environment.
l With auto sleep, auto shutdown and other power saving functions as well as battery rest capacity indicating function. Smart portable design with high reliability, it can anti vibration, shock and electromagnetic interference, suitable for harsh operating environment.
To make sure the probe working properly, it needs to use couplant to isolate the air between the probe surface and the measured workpiece surface. The liquid used for the coupling between the probe and workpiece is called as couplant.
Transducer Selection
Model |
Freq |
Diam |
Measuring Range |
Lower limit |
Description |
N05 |
5MHz |
10mm |
1.0mm~600mm (in steel) |
Φ20mm×3.0mm |
Normal measurement |
N05/90° |
5MHz |
10mm |
1.0mm~600mm (in steel) |
Φ20mm×3.0mm |
Normal measurement |
N07 |
7MHz |
6mm |
1.0mm~600mm (in steel) |
Φ15mm×2.0mm |
For thin pipe wall or small curvature pipe wall measurement |
HT5 |
5MHz |
12mm |
1.0mm~600mm (in steel) |
30mm |
For high temperature (lower than 300℃) measurement |